• Operating mode: Semi-automatic
• Waveform: e-cube biphastic (truncated exponential type); impedance compensated.
• Energy: 200J fixed
• Adult:
- nominal 150 J into a 50 ohm load.
- Infant/child: nominal 50 joules into a 50 ohm load.
• Shock energy accuracy: within ± 3% (200J into 50 ohm
• Shock to shock cycle time: Typically less than 20 seconds.
• Protocol: Voice prompts and indicators guide through protocol.
• Charge control: Automatic by Software (Arrhytmia Detection System and Charging control).
• Charge time from "Shock advised": <10 seconds, typical
• Voice instruction: Voice messages guide through use of defibrillator
• Controls: Shock Button, i-Button, On/Off Button.
• Indicators: 7 LEDs (different colors), i-Button.
• Patient isolation: Type BF.
• Battery: 12 Volt DC, 4.2 Ah, lithium manganese dioxide, disposable long-life primary cell
• Battery capacity: Min.120 Shocks or 4 hours of operating time.
• Data recording and transmission: Data stored: First 40 minutes of ECG and the entire incident's events and analysis decision.
• Size: 220x 281 x 82 mm
• Weight: 2.2 Kg (4.48 lbs) with battery pack.
• Environmental:
• Temperature: from 32 °F to 110 °F (from 0 °C to 40° C)
• Humidity: from 5 % to 95 % (non-condensing)
• Storage Conditions:
• Temperature: from 32 °F to 110 °F (from 0 °C to 40° C)
• Humidity: from 5 % to 95 % (non-condensing)
• Mechanical tests (shock drop tolerance): Meets IEC 60601-1 clause 21 (Mechanical Strenght)
• Vibration: Meets EN 1789 random and swept sine, road ambulance specification in operating and standby states.
• Impermeability: Meets IEC 60529: IP54
• ESD (Electrostatic discharge): Meets
• IEC 61000-4-2:2001
• EMI (radiated) meets: IEC 60601-1-2 limits, method EN 55011:1998+ A1:1999 +A2:2002, Group 1, Class B
• EMI (immunity) meets: IEC 60601-1-2 limits, method EN 61000-4-3: 2001 Level 3 (10V/m 80MHz to 2500MHz)
|
Manual |
|
CE certificate |
|