Facebook
  My Account
0
My cart
Defibrillators
[+] View All

Shipped in 10 days
I-PAD PUBLIC ACCESS DEFIBRILLATOR - french

The i-PAD NF 1200 is a semi-automatic external defibrillator designed for minimally trained individuals. It provides simple and direct voice prompts and indications for a straightforward rescue operation. It is lightweight and battery powered for maximum portability. It is designed to treat Ventricular Fibrillation (VF) and Fast Ventricular Tachycardia. These two are the most common causes of sudden cardiac arrest (SCA). In SCA, the heart of the victim suddenly stops pumping. The only effective treatment for VF is the application of a defibrillating shock. It may be used on children 1-8 years old (less than 25 kg) with pediatric reduced energy pads, over with adult pads.


Item #: 67831
Unità di vendita: 1 pc.
Quantitativi minimi: 1
NSIS:
CND:

 
€ 1.690,00
(€ 2.061,80 IVA Inclusa 22%)


• Operating mode: Semi-automatic
• Waveform: e-cube biphastic (truncated exponential type); impedance compensated.
• Energy: 200J fixed
• Adult:
- nominal 150 J into a 50 ohm load.
- Infant/child: nominal 50 joules into a 50 ohm load.
• Shock energy accuracy: within ± 3% (200J into 50 ohm
• Shock to shock cycle time: Typically less than 20 seconds.
• Protocol: Voice prompts and indicators guide through protocol.
• Charge control: Automatic by Software (Arrhytmia Detection System and Charging control).
• Charge time from "Shock advised": <10 seconds, typical
• Voice instruction: Voice messages guide through use of defibrillator
• Controls: Shock Button, i-Button, On/Off Button.
• Indicators: 7 LEDs (different colors), i-Button.
• Patient isolation: Type BF.
• Battery: 12 Volt DC, 4.2 Ah, lithium manganese dioxide, disposable long-life primary cell
• Battery capacity: Min.120 Shocks or 4 hours of operating time.
• Data recording and transmission: Data stored: First 40 minutes of ECG and the entire incident's events and analysis decision.
• Size: 220x 281 x 82 mm
• Weight: 2.2 Kg (4.48 lbs) with battery pack.

• Environmental:

• Temperature: from 32 °F to 110 °F (from 0 °C to 40° C)
• Humidity: from 5 % to 95 % (non-condensing)
• Storage Conditions:
• Temperature: from 32 °F to 110 °F (from 0 °C to 40° C)
• Humidity: from 5 % to 95 % (non-condensing)

• Mechanical tests (shock drop tolerance): Meets IEC 60601-1 clause 21 (Mechanical Strenght)
• Vibration: Meets EN 1789 random and swept sine, road ambulance specification in operating and standby states.
• Impermeability: Meets IEC 60529: IP54
• ESD (Electrostatic discharge): Meets
• IEC 61000-4-2:2001
• EMI (radiated) meets: IEC 60601-1-2 limits, method EN 55011:1998+ A1:1999 +A2:2002, Group 1, Class B
• EMI (immunity) meets: IEC 60601-1-2 limits, method EN 61000-4-3: 2001 Level 3 (10V/m 80MHz to 2500MHz)